Skip Navigation


Japanese Journal of Clinical Oncology Advance Access originally published online on May 30, 2007
Japanese Journal of Clinical Oncology 2007 37(5):391-398; doi:10.1093/jjco/hym032
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
37/5/391    most recent
hym032v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Request Permissions
Google Scholar
Right arrow Articles by Huh, H.
Right arrow Articles by Kim, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Huh, H.
Right arrow Articles by Kim, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?


© 2007 Foundation for Promotion of Cancer Research

Rectum Dose Analysis Employing a Multi-purpose Brachytherapy Phantom

HyunDo Huh1,2, WooChul Kim1, John J K Loh1, Suk Lee3,, Chul Yong Kim3, SangHoon Lee2, DongOh Shin4, Dongho Shin5, SamJu Cho2, Jisun Jang2, Sangwook Lim2, Kwang Hwan Cho2, Sooil Kwon2 and SeongHoon Kim6

1 Department of Radiation Oncology, College of Medicine, Inha University
2 Department of Medical Physics, Kyonggi University
3 Department of Radiation Oncology, College of Medicine, Korea University
4 Department of Radiation Oncology, Medicine, College of Medicine, KyungHee University
5 Research Institute and Hospital, National Cancer Center
6 Department of Radiation Oncology, College of Medicine, Hanyang University, Seoul, Korea

For reprints and all correspondence: Suk Lee, Department of Radiation Oncology, College of Medicine, Korea University, 5 Ga Anam-dong, Sungbuk-gu, Seoul 136-701, Korea. E-mail: sukmp{at}hanmail.net

Received October 3, 2006; accepted December 25, 2006

Purpose: It is difficult to reproduce a brachytherapy measurement because of changes in the rectal shape during inter-fraction. We constructed a multi-purpose brachytherapy phantom (MPBP) and reproduced the same conditions found in actual therapy. We further attempted to apply the measured optimal dose to reduce rectal complications.

Methods: A measured dose was administered at rectal reference point R1 using a diode detector in four patients who used a tandem and ovoid in brachytherapy for carcinoma of the cervix. A total number of 20 rectal dose measurements were performed five times per patient. In addition, discrepancies in the set-up of the diode detector were analyzed with each repetitive measurement. After reproducing the same conditions as found in actual therapy using a multi-function applicator (MFA) in the multi-purpose brachytherapy phantom constructed for this study, the dose was measured at reference points in the rectum using a thermoluminescence dosimeter (TLD).

Results: According to the discrepancies measured in the set-up using a diode detector, Patient 1 showed a maximum value of 11.25 ± 0.95 mm in the Y direction, Patients 2 and 3 exhibited 9.90 ± 2.40 mm and 20.85 ± 4.50 mm in the Z direction, respectively. Patient 4 showed 19.15 ± 3.33 mm in the Z direction. In addition, values of the mean dose according to the position of the diode detector were recorded as 122.82 ± 7.96–323.78 ± 11.16 cGy. In the measured results for TLD in an MPBP, relative error for Patients 1 and 4 at the rectal reference point R2 were a maximum of 8.6 and 7.7%, respectively. For Patients 2 and 3 they were 1.7 and 1.2%, respectively. Furthermore, the dose measured at point R1 and R2 exhibited values approximately 1.7–8.6% higher than the dose calculated in advance, excluding point R1 in Patient 2. The discrepancies in the set-up owing to repetitive measurements and alterations in dosage according to these changes were not analyzed. It was evident that the relative error between the calculated and measured value was within 15%, which was allowable according to the recommendations by the American Association of Physicists in Medicine (AAPM).

Conclusions: The multi-purpose brachytherapy phantom constructed for this study successfully reproduced an optimal dose measured under the same conditions found in actual therapy in which the dose was precisely analyzed at a rectal reference point. In addition, these results were considered reliable and applicable for dose optimization before applying therapy using the measured data from the phantom in order to reduce rectal complications.

Key Words: brachytherapy • multi-purpose brachytherapy phantom (MPBP) • MFA • diode detector • TLD


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.